64- Versus 16-Slice CT Angiography for Coronary Artery Stent Assessment
- 1 January 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 41 (1) , 22-27
- https://doi.org/10.1097/01.rli.0000191365.88345.e1
Abstract
We sought to assess the visualization of different coronary artery stents and the delineation of in-stent stenoses using 64- and 16-slice multidector computed tomography (MDCT). A total of 15 different coronary stents with a simulated in-stent stenosis were placed in a vascular phantom and scanned with a 16-slice and a 64-slice MDCT at orientations of 0 degree, 45 degrees, and 90 degrees relative to the scanner's z-axis. Visible lumen diameter and attenuation in the stented and the unstented segment of the phantom were measured. Three readers assessed stenosis delineation and visualization of the residual lumen using a 5-point scale. Artificial lumen narrowing (ALN) was significantly reduced with 64-slice CT compared with 16-slice CT. At an angle of 0 degree, 45 degrees, and 90 degrees relative to the scanner's z-axis, the ALN for 16-slice CT was 42.2%, 39.8%, and 44.0% using a slice-thickness of 1.0 mm and 40.9%, 40.4%, and 41.6% using a slice thickness of 0.75 mm, respectively. With 64-slice CT, the ALN was 39.1%, 37.3%, and 36.0% at the respective angles. The differences between attenuation values in the stented and unstented segment of the tube were significantly lower for 64-slice CT. Mean visibility scores were significantly higher for 64-slice CT. Use of the 64-slice CT results in superior visualization of the stent lumen and in-stent stenosis compared with 16-slice CT, especially when the stent is orientated parallel to the x-ray beam.Keywords
This publication has 14 references indexed in Scilit:
- Noninvasive assessment of left main coronary stent patency with 16-slice computed tomographyThe American Journal of Cardiology, 2005
- Performance Evaluation of a 64-Slice CT System with z-Flying Focal SpotRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2004
- Coronary Artery Stents in Multislice Computed TomographyInvestigative Radiology, 2004
- Assessment Of Coronary Arterial Stents By Multislice-Ct AngiographyActa Radiologica, 2003
- Sirolimus-Eluting Stents versus Standard Stents in Patients with Stenosis in a Native Coronary ArteryNew England Journal of Medicine, 2003
- Multislice spiral computed tomography for the detection of coronary stent restenosis and patencyInternational Journal of Cardiology, 2003
- Imaging of coronary artery stents using multislice computed tomography: in vitro evaluationEuropean Radiology, 2003
- Detection of Coronary Artery Stenoses With Thin-Slice Multi-Detector Row Spiral Computed Tomography and Multiplanar ReconstructionCirculation, 2003
- Reliable Noninvasive Coronary Angiography With Fast Submillimeter Multislice Spiral Computed TomographyCirculation, 2002
- Multislice Spiral Computed Tomography for In-Stent RestenosisCirculation, 2002