Validation of oxygen consumption measurements during artificial ventilation
- 1 November 1989
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 67 (5) , 2129-2134
- https://doi.org/10.1152/jappl.1989.67.5.2129
Abstract
We describe a system for the absolute calibration of indirect calorimeters, under the conditions of artificial ventilation and increased inspired O2 concentration, in which butane, at a measured flow rate, is burned downstream of an artificial lung. One milliliter of butane requires 6.4 ml O2 for its combustion, and the respiratory quotient is 0.615. With the closed-circuit O2-replenishment method there was no significant systematic error in the measurement of either O2 consumption or CO2 output and a random error with a SD of 8.3 ml/min for O2 consumption and 6.3 ml/min for CO2 output. There were no significant differences in the errors with inspired O2 concentrations between 23.8 and 59.5% and O2 consumptions between 89 and 366 ml/min.This publication has 5 references indexed in Scilit:
- The Accuracy and Precision of an Open-Circuit System To Measure Oxygen Consumption and Carbon Dioxide Production in NeonatesPediatric Research, 1986
- A COMPUTER-CONTROLLED INDIRECT CALORIMETER FOR THE MEASUREMENT OF ENERGY-EXPENDITURE IN ONE OR 2 SUBJECTS SIMULTANEOUSLY1986
- A SYSTEM FOR THE CONTINUOUS MEASUREMENT OF OXYGEN UPTAKE AND CARBON DIOXIDE OUTPUT IN ARTIFICIALLY VENTILATED PATIENTSBritish Journal of Anaesthesia, 1983
- A Systematic Method for Validation of Gas Exchange MeasurementsAnesthesiology, 1982
- Analysis of error in the determination of respiratory gas exchange at varying FIO2Journal of Applied Physiology, 1981