In Vitro Testing of Artificial Heart Valves: Comparison Between Newtonian and Non‐Newtonian Fluids
- 1 January 1996
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 20 (1) , 37-46
- https://doi.org/10.1111/j.1525-1594.1996.tb04416.x
Abstract
The in vitro testing of artificial heart valves is often performed with simple fluids like glycerol solutions. Blood, however, is a non‐Newtonian fluid with a complex viscoelastic behavior, and different flow fields in comparable geometries may result. Therefore, we used different polymer solutions (Polyacrylamid, Xanthan gum) with blood‐like rheological properties as well as various Newtonian fluids (water, glycerol solutions) in our heart valve test device. Hydrodynamic parameters of Björk‐Shiley heart valves with a tissue annulus diameter (TAD) of 21–29 mm were investigated under aortic flow conditions. Major results can be summarized as follows. The mean systolic pressure differences depend on the model fluids tested. Closing time and closing volume are not influenced by the rheological behavior of fluids. These parameters depend on TAD and the pressure differences across the valve. In contrast, rheological behavior has a pronounced influence upon leakage flow and leakage volume, respectively. Results show furthermore that the apparent viscosity data as a function of shear rate are not sufficient to characterize the rheological fluid behavior relevant to hydrodynamic parameters of the heart valves investigated. Therefore, similarity in the yield curves of non‐Newtonian test fluids mimicing blood is only a prerequisite for a suitable test fluid. More information about the viscous and elastic component of the fluid viscosity is required, especially in geometries where a complex flow field exists as in the case of leakage flow.Keywords
This publication has 20 references indexed in Scilit:
- Testung hydrodynamischer Eigenschaften von Herzklappenprothesen mit einem neuen Prüfstand - Testing of the Hydrodynamic Properties of Artificial Heart Valves with a New Test ApparatusBiomedizinische Technik/Biomedical Engineering, 1994
- Viscoelastic behaviour of human blood and polyacrylamide model fluids for heart valve testingJournal de Physique III, 1993
- Hydrodynamischer In-vitro-Vergleich einer neuen künstlichen Herzklappe mit der Björk-Shiley-Klappe (Standard) - Hydrodynamic In-vitro Comparison of a New Artificial Heart Valve with the Björk-Shiley Valve (Standard)Biomedizinische Technik/Biomedical Engineering, 1991
- In-vitro wall shear measurements at aortic valve prosthesesJournal of Biomechanics, 1984
- In-Vitro Fluid Dynamic Characteristics of the Abiomed Trileaflet Heart Valve ProsthesisJournal of Biomechanical Engineering, 1983
- The Björk-Shiley and Ionescu-Shiley Heart Valve ProsthesesIn Vitro Comparison of Their Hydrodynamic Performance in the Mitral PositionScandinavian Journal of Thoracic and Cardiovascular Surgery, 1983
- Regurgitation of prosthetic heart valves: Dependence on heart rate and cardiac outputThe American Journal of Cardiology, 1983
- Flow Pattern and Velocity Field Distal to Human Aortic and Artificial Heart Valves as Measured Simultaneously by Ultramicroscope Anemometry in Cylindrical Glass TubesThe Thoracic and Cardiovascular Surgeon, 1980
- IN VITRO COMPARISON BETWEEN THE HYDRODYNAMIC CHARACTERISTICS OF THE HANCOCK 250 (MODIFIED ORIFICE) XENOGRAFTS AND THE BJÖRK-SHILEY AORTIC VALVE PROSTHESESAsaio Journal, 1977
- Regurgitation and Haemolysis in Artificial Heart Valves: An Experimental Study of Overlapping and Non-overlapping Closing Mechanisms and of Paraprosthetic LeakageScandinavian Journal of Thoracic and Cardiovascular Surgery, 1974