Tonometry in the general practice setting (I)
- 27 May 2009
- journal article
- clinical trial
- Published by Wiley in Acta Ophthalmologica
- Vol. 71 (1) , 103-108
- https://doi.org/10.1111/j.1755-3768.1993.tb04970.x
Abstract
Intraocular pressure was measured in 132 consecutive patients over the age of 20 in a general practice with two general practitioners (GP). Tono-Pen tonometry was compared to Goldmann applantation tonometry in a masked, randomized design. Also, agreement between Tono-Pen measurements by the GPs and the ophthalmologist was estimated. Both mean difference between Tono-Pen and Goldmann applantation measurements and mean differences between Tono-Pen readings made by GPs and ophthalmologist were less than 1 mmHg. Ninety-five percent limits of agreement between the two methods were approximately -4.5 mmHg and 5.5 mmHg, and between observers approximately -4 mmHg and 5 mmHg. The Tono-Pen was agreeable both to patients and physicians. The GPs made as accurate measurements as the ophthalmologist. The wide limits of agreement are clinically much more important than the small differences between means. Random variation was much greater than systematic variation. Wide limits of agreement will apply to all known methods of clinical tonometry.Keywords
This publication has 14 references indexed in Scilit:
- The Effect of Perkins, Tono-Pen, and Schiötz Tonometry on Intraocular PressureAmerican Journal of Ophthalmology, 1991
- Evaluation of the Tono-Pen and the Pulsair TonometersAmerican Journal of Ophthalmology, 1990
- An Evaluation of the Tono-Pen for the Measurement of Diurnal Intraocular PressureAmerican Journal of Ophthalmology, 1989
- The Tono-PenArchives of Ophthalmology (1950), 1988
- Comparison of Intraocular Pressure Measurements With the Oculab Tono-Pen vs Manometry in Humans Shortly After DeathAmerican Journal of Ophthalmology, 1988
- Comparison of the Tono-Pen to the Goldmann Applanation TonometerArchives of Ophthalmology (1950), 1988
- Clinical Comparison of the Oculab Tono-Pen to the Goldmann Applanation TonometerOphthalmology, 1987
- Clinical Evaluation of the Oculab Tono-PenAmerican Journal of Ophthalmology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Repeated Applanation TonometryAmerican Journal of Ophthalmology, 1968