Intercomparison of whole-body counters using a multinuclide calibration phantom
- 1 February 1991
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 36 (2) , 191-198
- https://doi.org/10.1088/0031-9155/36/2/004
Abstract
Whole-body counters in the UK have been compared using a multinuclide anthropomorphic phantom. A standard Bush phantom was modified by inserting channels into the long axis of each section. Radionuclide sources sealed in a urea-formaldehyde polymer were then inserted into the channels to simulate distributions of radioactivity in a human. The phantom was taken to 10 whole-body counters in the UK and estimates of 134Cs, 137Cs and 40K were obtained both separately and as mixtures. Results showed close agreement between the median estimates and the known activities. The technique also allowed diagnosis of particular problems in calibration for several of the counters.Keywords