Reference Intervals for a Complete Blood Count Determined on different Automated Haematology Analysers: Abx Pentra 120 Retic, Coulter Gen-S, Sysmex SE 9500, Abbott Cell Dyn 4000 and Bayer Advia 120
- 29 January 2002
- journal article
- research article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 40 (1)
- https://doi.org/10.1515/cclm.2002.014
Abstract
We processed 317 samples from healthy adult volunteers for a complete blood count, including leukocyte differentials and reticulocyte parameters, through five new-generation haematology analysers: Abx Pentra 120 Retic, Coulter Gen-S, Sysmex SE 9500, Abbott Cell Dyn 4000 and Bayer Advia 120. From these data non-parametric 2.5-97.5 percentile reference intervals were calculated for all parameters on all analysers. Some differences were found compared with previously reported reference intervals. Reference intervals for platelet parameters and reticulocytes agreed with these usually accepted. For red blood cell parameters, including haemoglobin and haematocrit, and white blood cell count, including absolute white blood cell differentials, our calculated reference intervals were in agreement with less frequently cited earlier reports, but were lower compared to the usually accepted reference intervals.Keywords
This publication has 14 references indexed in Scilit:
- The White Blood Cell Differential: Three Methods Comparedcclm, 2001
- Flow Cytometric Reticulocyte CountingAmerican Journal of Clinical Pathology, 2001
- The determination of a reference range for new platelet parameters produced by the Bayer ADVIATM 120 full blood count analyserClinical and Laboratory Haematology, 2000
- Evaluation of the Abbott CELL-DYN 4000 Hematology AnalyzerAmerican Journal of Clinical Pathology, 2000
- Reticulocyte Cellular Indices: A New Approach in the Diagnosis of Anemias and Monitoring of Erythropoietic FunctionCritical Reviews in Clinical Laboratory Sciences, 2000
- Automated Reticulocyte Counting and Immature Reticulocyte Fraction Measurement: Comparison of ABX PENTRA 120 Retic, Sysmex R-2000, Flow Cytometry, and Manual CountsAmerican Journal of Clinical Pathology, 1999
- Preliminary Evaluation of the New Hematology Analyzer COULTER® GEN•S™ in a University Hospitalcclm, 1999
- Normal Reference Laboratory ValuesNew England Journal of Medicine, 1998
- The routine measurement of platelet volume: a comparison of aperture-impedance and flow cytometric systemsClinical and Laboratory Haematology, 1985
- Haematologic Profile of Natural Populations: Red Cell ParametersBritish Journal of Haematology, 1977