Evaluation of the COBAS TaqMan HCV Test with Automated Sample Processing Using the MagNA Pure LC Instrument
- 1 January 2005
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 43 (1) , 293-298
- https://doi.org/10.1128/jcm.43.1.293-298.2005
Abstract
The COBAS TaqMan HCV Test (TaqMan HCV; Roche Molecular Systems Inc., Branchburg, N.J.) for hepatitis C virus (HCV) performed on the COBAS TaqMan 48 Analyzer (Roche Molecular Systems) currently relies on a manual sample processing method. Implementation of an automated sample processing method would facilitate the clinical use of this test. In this study, we evaluated the performance characteristics of TaqMan HCV following automated sample processing by the MagNA Pure LC instrument (MP; Roche Applied Science, Indianapolis, Ind.). The analytical sensitivity of TaqMan HCV following sample processing by MP was 8.1 IU/ml (95% confidence interval, 6.1 to 15.2). The assay showed good linearity ( R 2 = 0.99) across a wide range of HCV RNA levels (25 to 5 × 10 6 IU/ml), with coefficients of variation ranging from 10% to 46%. Among 83 clinical specimens, the sensitivity and specificity of TaqMan HCV were 100% and 95%, respectively, when compared to the COBAS AMPLICOR hepatitis C virus test, version 2.0 (COBAS AMPLICOR; Roche Molecular Systems), with TaqMan HCV detecting two more HCV RNA-positive specimens than COBAS AMPLICOR. Both specimens were confirmed to be HCV RNA positive by the VERSANT HCV RNA qualitative test (Bayer HealthCare LLC, Tarrytown, N.Y.). There was also strong correlation ( R 2 = 0.95) and good agreement between the results from TaqMan HCV and the VERSANT HCV RNA 3.0 assay (bDNA) (Bayer HealthCare LLC) among a group of 93 clinical specimens. The MP is a versatile, labor-saving sample processing platform suitable for reliable performance of TaqMan HCV.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of the MagNA Pure LC Instrument for Extraction of Hepatitis C Virus RNA for the COBAS AMPLICOR Hepatitis C Virus Test, Version 2.0Journal of Clinical Microbiology, 2003
- Comparison of the VERSANT™ HCV RNA qualitative assay (transcription-mediated amplification) and the COBAS AMPLICOR™ hepatitis C virus test, version 2.0, in patients undergoing interferon-ribavirin therapyDiagnostic Microbiology and Infectious Disease, 2003
- Multicenter Evaluation of the VERSANT HCV RNA Qualitative Assay for Detection of Hepatitis C Virus RNAJournal of Clinical Microbiology, 2003
- Sensitivity and reproducibility of HCV quantitation in chimpanzee sera using TaqMan real-time PCR assayJournal of Virological Methods, 2002
- Assessment, by Transcription-Mediated Amplification, of Virologic Response in Patients with Chronic Hepatitis C Virus Treated with Peginterferon α-2aJournal of Clinical Microbiology, 2001
- Detection of residual hepatitis C virus RNA by transcription-mediated amplification in patients with complete virologic response according to polymerase chain reaction-based assaysHepatology, 2000
- Performance Characteristics of a Quantitative, Homogeneous TaqMan RT-PCR Test for HCV RNAThe Journal of Molecular Diagnostics, 2000
- The past incidence of hepatitis C virus infection: Implications for the future burden of chronic liver disease in the United StatesHepatology, 2000
- The Prevalence of Hepatitis C Virus Infection in the United States, 1988 through 1994New England Journal of Medicine, 1999
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986