Flat-response, subkiloelectronvolt x-ray detector with a subnanosecond time response
- 1 August 1978
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 49 (8) , 1204-1205
- https://doi.org/10.1063/1.1135548
Abstract
A detector that has a flat response to x rays with energies between 100 eV and 1.5 keV has been built. It has been used to measure x rays emitted from gold disks irradiated by the Argus laser at Lawrence Livermore Laboratory. Results from this detector are in good agreement with results from other techniques.Keywords
This publication has 1 reference indexed in Scilit: