Prevalence of nontuberculous mycobacteria in water supplies of hemodialysis centers
- 1 December 1988
- journal article
- research article
- Published by American Society for Microbiology in Applied and Environmental Microbiology
- Vol. 54 (12) , 3122-3125
- https://doi.org/10.1128/aem.54.12.3122-3125.1988
Abstract
Infection of hemodialysis patients with nontuberculous mycobacteria (NTM) has been associated with water used in reprocessing hemodialyzers. This study was conducted to determine the prevalence of NTM and other bacteria in water samples collected over a 13-week period from 115 randomly selected dialysis centers in the United States. Total viable counts were determined by membrane filter assays; increased recovery of NTM was obtained by dosing a portion of each water sample with 1% formaldehyde (HCHO) before filtering. NTM were widely distributed and occurred with a high frequency in water supplies in dialysis centers. NTM were detected in water from 95 centers (83%), and 50% of all samples examined contained NTM. The results of this study support recommendations to use 4% HCHO or a chemical germicidal equivalent for disinfecting dialyzers that are to be reused.This publication has 17 references indexed in Scilit:
- Infections with Mycobacterium chelonei in Patients Receiving Dialysis and Using Processed HemodialyzersThe Journal of Infectious Diseases, 1985
- Hemodialyzer reuse: practices in the United States and implication for infection control.1985
- Mycobacteria in waterJournal of Applied Bacteriology, 1984
- Bacterial contamination of the blood compartment originating from the dialysate in haemodialysersJournal of Hospital Infection, 1984
- Spectrum of Disease Due to Rapidly Growing MycobacteriaClinical Infectious Diseases, 1983
- Peritonitis Due to a Mycobacterium chelonei-Like Organism Associated with Intermittent Chronic Peritoneal DialysisThe Journal of Infectious Diseases, 1982
- Disseminated infection with Mycobacterium chelonei in a haemodialysis patientTubercle, 1981
- Bacteriological Contamination of DialyzersNephron, 1978
- Factors that influence microbial contamination of fluids associated with hemodialysis machines.1974
- PROBABLE MYCOBACTERIUM FORTUITUM SEPTICÆMIA: COMPLICATION OF HOME DIALYSISThe Lancet, 1971