Massive arterial air embolism during cardiopulmonary bypass: antegrade blood cardioplegia delivered by the pump — an accident waiting to happen
- 1 March 1996
- journal article
- case report
- Published by SAGE Publications in Perfusion
- Vol. 11 (2) , 157-161
- https://doi.org/10.1177/026765919601100209
Abstract
Of the many possible causes of air embolism occurring in patients undergoing cardiopulmonary bypass (CPB), human error due to the perfusionist or the surgeon accounts for the vast majority. This case, however, presents a previously unreported, but recognized, cause of air embolism, due to a technical problem encountered during the administration of blood cardioplegia. The nature of some of the older CPB pumps allows the administration of cardioplegia at a time when the main CPB pump is not rotating. A situation may then arise whereby air may be entrained and delivered to the patient. The management of massive air embolism is discussed, and recommendations are made to prevent such an occurrence happening in the future.Keywords
This publication has 12 references indexed in Scilit:
- A simple, safe and economical method for administration of continuous warm blood cardioplegiaPerfusion, 1994
- Hyperbaric oxygenation for arterial air embolism during cardiopulmonary bypassThe Annals of Thoracic Surgery, 1993
- Retrograde cerebral perfusion against massive air embolism during cardiopulmonary bypassThe Journal of Thoracic and Cardiovascular Surgery, 1992
- Primary myocardial revascularizationThe Journal of Thoracic and Cardiovascular Surgery, 1984
- Management of Massive Air Embolism during Open-heart Surgery with Retrograde Perfusion of the Cerebral Vessels and Hyperbaric OxygenationThe Thoracic and Cardiovascular Surgeon, 1983
- Brain damage during open-heart surgery.Thorax, 1982
- Etiologic Factors in Neuropsychiatric Complications Associated with Cardiopulmonary BypassAnesthesia & Analgesia, 1982
- Massive air embolism during cardiopulmonary bypassThe Journal of Thoracic and Cardiovascular Surgery, 1980
- Air Embolism and Other Accidents Using Pump OxygenatorsThe Annals of Thoracic Surgery, 1980
- Runaway pump head: New cause of gas embolism during cardiopulmonary bypassThe Journal of Thoracic and Cardiovascular Surgery, 1979