Camino(R) intracranial pressure monitor: prospective study of accuracy and complications
Open Access
- 1 July 2000
- journal article
- research article
- Published by BMJ in Journal of Neurology, Neurosurgery & Psychiatry
- Vol. 69 (1) , 82-86
- https://doi.org/10.1136/jnnp.69.1.82
Abstract
OBJECTIVES The fibreoptic device is a type of intracranial pressure monitor which seems to offer certain advantages over conventional monitoring systems. This study was undertaken to analyse the accuracy, drift characteristics, and complications of the Camino® fibreoptic device. METHODS One hundred and eight Camino® intracranial pressure (ICP) devices, in their three modalities, were implanted during 1997. The most frequent indication for monitoring was severe head injury due to road traffic accidents. RESULTS Sixty eight probe tips were cultured; 13.2% of the cases had a positive culture without clinical signs of infection, and 2.9% had a positive culture with clinical signs of ventriculitis. The most common isolated pathogen wasStaphylococcus epidermidis. All patients were under cephalosporin prophylaxis during monitoring. Haemorrhage rate in patients without coagulation disorders was 2.1% and 15.3% in patients with coagulation abnormalities. Drift characteristics were studied in 56 cases; there was no drifting from the values expected according to the manufacturer's specifications in 34 probes. There was no relation between direction of the drift and duration of placement, nor between drift and time. CONCLUSIONS Although the complication and drift rates were similar to those reported elsewhere, there was no correlation between the direction of the drift and long term monitoring despite the fact that some published papers refer to overestimation of values with time with this type of device.Keywords
This publication has 25 references indexed in Scilit:
- Prospective study of zero drift in fiberoptic pressure monitors used in clinical practiceJournal of Neurosurgery, 1997
- Johann Jakob Wepfer (1620-95) and cerebral haemorrhage.Journal of Neurology, Neurosurgery & Psychiatry, 1997
- Systematic overestimation of intracranial pressure measured using a Camino pressure monitor.Journal of Neurology, Neurosurgery & Psychiatry, 1996
- The fiberoptic intraparenchymal cerebral pressure monitor in 244 patientsSurgical Neurology, 1996
- The safety and value of extradural intracranial pressure monitors in fulminant hepatic failureJournal of Hepatology, 1993
- Complications of intracranial pressure monitoring in fulminant hepatic failureThe Lancet, 1993
- Evaluation of a fiberoptic intracranial pressure monitorJournal of Neurosurgery, 1990
- Complications of intracranial pressure monitoring in trauma patientsNeurosurgery, 1989
- Continuous monitoring of intracranial pressure with a miniaturized fiberoptic deviceJournal of Neurosurgery, 1987
- Ventriculostomy-Related InfectionsNew England Journal of Medicine, 1984