Determination of Retinol-Binding Protein in Serum by Kinetic Immunonephelometry with Polyethylene Glycol Pretreatment

Abstract
This work describes the use of polyethylene glycol as a pretreatment reagent to remove endogenous light scattering material from serum samples prior to automated immunonephelometric analysis on a centrifugal analyser. An assay system for retinol-binding protein is described, which allows rapid (10 minutes) quantitation of retinol-binding protein in serum samples with a detection limit of 5 mg/L and between-assay coefficients of variation ranging from 2·9% to 4·0%. The assay range is 5–80 mg/L and accuracy comparisons with a Mancini single radial immunodiffusion method yield a regression line y=0·89x+0·52 ( r=0·98, n=22). The problem of analyte precipitation associated with use of pretreatment regimes is discussed.