Clinical Evaluation of a Prototype Motion Artifact Resistant Pulse Oximeter in the Recovery Room
- 1 August 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 83 (2) , 269-272
- https://doi.org/10.1097/00000539-199608000-00012
Abstract
The frequency and nature of spurious pulse oximetry readings were compared using both a conventional pulse oximeter (CPO) and a prototype Masimo signal extraction technology pulse oximeter (Masimo SET). At a university hospital, 50 ASA physical status I-IV adult patients who underwent general or spinal-epidural anesthesia were selected from a group of 250 patients on the basis of high-alarm generation with routine postoperative pulse oximetry. Pulse oximetry data were recorded simultaneously from both devices with a computer. Overall, the CPO alarm frequency (i.e., oxygen saturation < 90%, or complete signal loss) was once every 13 min, and 87% of these alarms were considered false. Alarms were considered false based on reference electrocardiographs (16 patients), arterial blood gases (7 patients), and clinical assessment. The prototype Masimo SET device alarm frequency was once every 30 min, and 59% of these were considered false. During arm motion with 15 patients, the CPO device produced spurious signals on 54 occasions compared with five for the prototype Masimo SET. The incidence of artifactual pulse oximetry events during patient motion appear to be substantially reduced with the prototype Masimo SET device, relative to a CPO device.Keywords
This publication has 7 references indexed in Scilit:
- DOES A PROTECTIVE SHEATH ALLOWING CLINICAL REUSE ALTER SENSITIVITY AND RESPONSE TIME OF DISPOSABLE PULSE OXIMETER PROBES?Anesthesiology, 1994
- Postanesthesia monitoring revisited: Frequency of true and false alarms from different monitoring devicesJournal of Clinical Anesthesia, 1994
- Recent Developments in Pulse OximetryAnesthesiology, 1992
- EFFECTS OF DEFAULT ALARM LIMITSETTINGS ON ALARM DISTRIBUTION IN TELEMETRIC PULSE OXIMETRY NETWORK IN WARD SETTINGAnesthesiology, 1991
- Pulse Oximeter Failure Thresholds in Hypotension and VasoconstrictionAnesthesiology, 1990
- Pulse OximetryAnesthesiology, 1989
- Blood Flow Limits and Pulse Oximeter Signal DetectionAnesthesiology, 1987