Assessment of image intensifier and distortion for DSA localization studies.
- 1 January 1997
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 70 (829) , 70-73
- https://doi.org/10.1259/bjr.70.829.9059298
Abstract
The authors present methods of correcting pincushion and S distortions of an image intensifier, and of measuring the geometrical parameters of the imaging device used for localization from digital subtraction angiography brain studies. Pincushion and S distortions of the image intensifier are corrected by a calibration grid. A test pattern is used to study effectiveness of the corrections. Intrinsic geometrical parameters (source-to-image distance, centre of X-ray projection) of the apparatus are measured by the use of a calibration phantom. Short-range and long-range time drift of the distortion, as well as influence of the parameters of image acquisition on the accuracy of the localization results, are considered. The results obtained successfully demonstrate the accuracy of the correction, provided that the apparatus is warm.Keywords
This publication has 10 references indexed in Scilit:
- Frameless method of stereotaxic localization with DSA.Radiology, 1993
- Correction of aberration in image-intensifier systemsComputerized Medical Imaging and Graphics, 1992
- Accurate characterization of image intensifier distortionMedical Physics, 1991
- Knowledge-based system for the three-dimensional reconstruction of blood vessels from two angiographic projectionsMedical & Biological Engineering & Computing, 1991
- A knowledge-based approach for 3-D reconstruction and labeling of vascular networks from biplane angiographic projectionsIEEE Transactions on Medical Imaging, 1991
- Analysis and correction of imperfections in the image intensifier–TV–digitizer imaging chainMedical Physics, 1991
- Digital TV tomography: Description and physical assessmentMedical Physics, 1990
- Real-time distortion correction of digital X-ray II/TV-systems: an application example for Digital Flashing Tomosynthesis (DFTS)The International Journal of Cardiovascular Imaging, 1990
- Image intensifier distortion correctionMedical Physics, 1987
- Digital subtraction angiography system evaluation with phantoms.Radiology, 1985