Assessment of response function in two PET scanners with and without interplane septa
- 1 January 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 10 (3) , 240-248
- https://doi.org/10.1109/42.97571
Abstract
The authors have assessed the response function both experimentally and theoretically for two commercial tomographs: CTI 931/08-12 and CTI 953B with and without interplane septa. Monte Carlo simulations were undertaken using the GEANT package from CERN. Spatial resolution (tomographic and axial) was calculated for line sources at various positions in the field of view. Sensitivity and scatter fraction (SF) were calculated for various source geometries as a function of energy discrimination. A very realistic response function in positron emission tomography (PET) is obtained by Monte Carlo methods, using global parameters to account for unsimulated phenomena such as scintillation light transport inside a detector block and its sharing among the various phototubes. Minor discrepancies remain for sensitivity and SF at high energy thresholds and may probably be explained by introducing the observed dispersion in the energy response for the various crystals within a detector block.Keywords
This publication has 32 references indexed in Scilit:
- Analytical study of the effect of collimation on the performance of PET cameras in 3-D imagingIEEE Transactions on Nuclear Science, 1990
- Toward the design of a positron volume imaging cameraIEEE Transactions on Nuclear Science, 1990
- Detector, shielding and geometric design factors for a high-resolution PET systemIEEE Transactions on Nuclear Science, 1990
- Improving the efficiency of emission tomography simulations using variance reduction techniquesIEEE Transactions on Nuclear Science, 1990
- Monte Carlo simulation of BaF/sub 2/ detectors used in time-of-flight positron emission tomographyIEEE Transactions on Nuclear Science, 1990
- Three dimensional reconstruction of PET data from a multi-ring cameraIEEE Transactions on Nuclear Science, 1989
- A study of the possibility of using multi-slice PET systems for 3D imagingIEEE Transactions on Nuclear Science, 1989
- Simulation studies for cylindrical positron tomographyIEEE Transactions on Nuclear Science, 1988
- The effect of collimation on scatter fraction in multi-slice PETIEEE Transactions on Nuclear Science, 1988
- Elektron-Photon-Kaskaden in BleiThe European Physical Journal A, 1965