A magnetic shielded room designed for magnetoencephalography
- 1 April 1989
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 60 (4) , 765-770
- https://doi.org/10.1063/1.1141016
Abstract
A magnetic shielded room has been constructed for magnetoencephalographic (MEG) measurements. The room was designed to match the noise levels of superconducting quantum interference device coupled second‐order gradiometers while minimizing the overall cost of the facility. The room measures 2.6×2.6×2.6 m3 internally and consists of a single high‐permeability alloy shell with an outer shell of ultrapure aluminum. The dc shielding factor at the center of the room is 200, and the ac shielding factor at 10 Hz is 2000. This is the frequency which has the maximum amplitude in a Fourier analysis of MEG results. The room is designed to operate without a door closure.Keywords
This publication has 7 references indexed in Scilit:
- Magnetic field fluctuations arising from thermal motion of electric charge in conductorsJournal of Applied Physics, 1984
- An eddy-current-shielded room with a partially closed entranceIl Nuovo Cimento D, 1983
- Aluminum shielded room for biomagnetic measurementsReview of Scientific Instruments, 1981
- The effect of shaking on magnetic shieldsIEEE Transactions on Magnetics, 1980
- SQUID instruments and shielding for low-level magnetic measurementsJournal of Applied Physics, 1977
- A Shielded Facility for Low-Level Magnetic MeasurementsJournal of Applied Physics, 1967
- ENHANCEMENT OF FERROMAGNETIC SHIELDING AGAINST LOW-FREQUENCY MAGNETIC FIELDSApplied Physics Letters, 1967