An inductively coupled plasma-time-of-flight mass spectrometer for elemental analysis. Part I: Optimization and characteristics
- 1 November 1994
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 5 (11) , 1008-1016
- https://doi.org/10.1016/1044-0305(94)80019-7
Abstract
An inductively coupled plasma-time-of-flight mass spectrometer (ICP-TOFMS) has been constructed and evaluated for elemental analysis. The instrument produces analog spectra similar to those from quadrupole inductively coupled plasma mass spectrometers. The large abundance of Ar ions is deflected away from the microchannel plate detector to reduce detector dead time and space-charge complications. The ICP-TOFMS, operated in a linear (nonreflecting) mode, currently has a resolving power of 500 (full width at half maximum). Present ion optics employed in the instrument require a trade-off between signal-to-noise ratio and resolving power. In addition, mass-dependent kinetic energies in the supersonic beam created in the ICP mass spectrometer interface cause a mass bias in the right-angle TOFMS because the ions must be steered to the detector to compensate for their velocity in the supersonic beam direction. In the current design the sampling duty cycle is only approximately 3%, thereby limiting sensitivity. However, positive potentials applied to the right-angle extraction region can increase sensitivity by a factor of 2–4 by slowing down the ions that enter the extraction zone. The transmission efficiency of the TOFMS is approximately 20% and is limited by divergence of the ion packet in the drift tube.Keywords
This publication has 13 references indexed in Scilit:
- Plasma source atmospheric pressure ionization detection of liquid injection using an ion trap storage/reflectron time-of-flight mass spectrometerAnalytical Chemistry, 1993
- Time-of-flight mass spectrometry with an electrospray ion beamAnalytical Chemistry, 1992
- Atmospheric pressure ionization time-of-flight mass spectrometry with a supersonic ion beamAnalytical Chemistry, 1991
- Large ion volume time-of-flight mass spectrometer with position- and velocity-sensitive detection capabilities for cluster beamsReview of Scientific Instruments, 1991
- A transversally and longitudinally focusing time-of-flight mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1991
- Fast-Pulse Detection for Isotopic Abundance Determination By Resonance Ionization, Time-Of-Flight Mass SpectrometryInstrumentation Science & Technology, 1988
- Quantification of pulsed ion currents produced in resonance ionization mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1987
- Ultrasonic nebulization of liquid samples for analytical inductively coupled plasma-atomic spectroscopy: an updateSpectrochimica Acta Part B: Atomic Spectroscopy, 1986
- Mass spectrometry of inductively coupled plasmasAnalytical Chemistry, 1986
- Isotopic analysis with the laser microprobe mass analyzerInternational Journal of Mass Spectrometry and Ion Processes, 1983