Electrical properties of implant encapsulation tissue
- 1 January 1994
- journal article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 22 (1) , 23-33
- https://doi.org/10.1007/bf02368219
Abstract
The purpose of this study was to determine the electrical properties of the encapsulation tissue that surrounds electrodes chronically implanted in the body. Two four-electrode arrays, fabricated from either epoxy or silicone rubber, were implanted in each of six adult cats for 82 to 156 days.In vivo measurements of tissue resistivity using the four-electrode technique indicated that formation of the encapsulation tissue resulted in a significant increase in the resistivity of the tissue around the arrays.In vitro measurements of tissue impedance using a four-electrode cell indicated that the resistivity of the encapsulation tissue was a function of the tissue morphology. The tight layers of fibroblasts and collagen that formed around the silicone rubber arrays had a resistivity of 627±108 Ω-cm (mean ± SD; n=6), which was independent of frequency from 10 Hz to 100 kHz, and was significantly larger than the resistivity of the epoxy encapsulation tissue at all frequencies between 20 Hz and 100 kHz. The combination of macrophages, foreign body giant cells, loose collagen, and fibroblasts that formed around the epoxy arrays had a frequency-dependent resistivity that decreased from 454±123 Ω-cm (n=5) to 193±98 Ω-cm between 10 Hz and 1 kHz, and was independent of frequency between 1 kHz and 100 kHz, with a mean value of 195 ±88 Ω-cm. The results indicate that the resistivity of the encapsulation tissue is sufficient to alter the shape and magnitude of the electric field generated by chronically implanted electrodes.Keywords
This publication has 35 references indexed in Scilit:
- Functional electrical stimulation of the latissimus dorsi muscle for use in cardiac assistIEEE Transactions on Biomedical Engineering, 1989
- Scarring around Cervical Epidural Stimulating ElectrodeNeurosurgery, 1983
- Effect of surface texture on the soft tissue response to polymer implantsJournal of Biomedical Materials Research, 1983
- The response to the intramuscular implantation of pure metalsBiomaterials, 1981
- Complications of dorsal column stimulationJournal of Neurosurgery, 1978
- Deleterious Effects of Prolonged Electrical Excitation of Striate Cortex in MacaquesBrain, Behavior and Evolution, 1977
- Tissue response to implanted polymers: The significance of sample shapeJournal of Biomedical Materials Research, 1976
- Dorsal column stimulation for control of painJournal of Neurosurgery, 1972
- The significance of implant shape in experimental testing of biological materials: Disc vs. RodJournal of Biomedical Materials Research, 1970
- Wound HealingScientific American, 1969