Abstract
A novel absorption detection method for highly multiplexed capillary electrophoresis is presented for zone electrophoresis and for micellar electrokinetic chromatography. The approach involves the use of a linear photodiode array on which a capillary array is imaged by a camera lens. Either a tungsten lamp or a mercury lamp can be used as the light source such that all common wavelengths for absorption detection are accessible by simply interchanging narrow-band filters. Each capillary spans several diodes in the photodiode array for absorption measurements. Over 100 densely packed capillaries can be monitored by a single photodiode array element with 1024 diodes. The detection limit for rhodamine 6G for each capillary in the multiplexed array is ∼1.8 × 10-8 M injected (S/N = 2). The cross-talk between adjacent capillaries is less than 0.2%. Simultaneous analysis of 96 samples is demonstrated.