Concomitant magnetic‐field‐induced artifacts in axial echo planar imaging
- 1 April 1998
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 39 (4) , 596-605
- https://doi.org/10.1002/mrm.1910390413
Abstract
When a linear magnetic field gradient is used, spatially higher‐order magnetic fields are produced to satisfy the Maxwell equations. It has been observed that the higher‐order magnetic field produced by the readout gradient causes axial echo planar images acquired with a horizontal solenoid magnet to shift along the phase‐encoding direction and lose image intensities. Both the shift and intensity reduction become increasingly severe as the slice offset from the isocenter increases. These phenomena are quantitatively analyzed, and good correlation between experiments and theory has been established. The analysis also predicts a previously unre‐ported Nyquist ghost on images with very large slice offsets. This ghost has been verified with computer simulations. Based on the analysis, several methods have been developed to eliminate the image shift, the intensity reduction, and the ghost. Selected methods have been implemented on a commercial scanner and proved effective in removing these image artifacts.Keywords
This publication has 4 references indexed in Scilit:
- Concomitant magnetic‐field‐induced artifacts in axial echo planar imagingMagnetic Resonance in Medicine, 1998
- Concomitant gradient terms in phase contrast MR: Analysis and correctionMagnetic Resonance in Medicine, 1998
- Nonaxial whole‐body instant imagingMagnetic Resonance in Medicine, 1993
- Concomitant magnetic field gradients and their effects on imaging at low magnetic field strengthsMagnetic Resonance Imaging, 1990