Reliability of the Reflotron in the determination of cholesterol
- 1 May 1990
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 50 (3) , 261-271
- https://doi.org/10.3109/00365519009091577
Abstract
Measurements of total cholesterol in the field by means of the Reflotron dry-chemistry system (capillary blood) were compared to total cholesterol obtained by a standardized conventional wet-chemistry method in a clinico-chemical laboratory (serum). A total of 1200 people participated in the study. Two identical Reflotron machines were used. In the first period of the study an excellent agreement was found between Reflotron measurements of a reference serum provided by the manufacturer (mean, 4.99 mmol/l; CV, 1.8%) and the stated value (4.97 mmol/l). In the rest of the study higher values and greater variation were found with the Reflotron (mean, 5.32 mmol/l CV 5.2%). Clearly the reflotron measurements in the latter period of study were not reliable. In the period with stable instruments most of the values obtained at the two Reflotron machines differed from each other by less than 10%, with a mean difference of 0.08 mmol/l. Reflotron (both machines) and wet-chemistry measurements agreed well for the first 500 participants in the study (mean difference, Reflotron-wet-chemistry, -0.008 mmol/l; 95% confidence interval, -0.035 to 0.019 mmol/l; correlation, 0.967). In this period most Reflotron values differed from wet-chemistry values by less than 9% below to 9% above. With the next 200 participants the Reflotron gave on average slightly higher values than wet-chemistry measurements. The coefficients of variation for measurement variation were higher for Reflotron than for wet-chemistry even in the period with stable instruments. In all parts of the study period a lower HDL-cholesterol level was associated with larger differences between total cholesterol determined by Reflotron and wet-chemistry.Keywords
This publication has 8 references indexed in Scilit:
- How accurately does the Reflotron dry‐chemistry system measure plasma total cholesterol levels when used as a community‐screening device?The Medical Journal of Australia, 1988
- An evaluation of the accuracy of the Reflotron system in the fieldThe Medical Journal of Australia, 1988
- Comparison of Reflotron and laboratory cholesterol measurementsThe Medical Journal of Australia, 1988
- Revised calibration of the Reflotron cholesterol assay evaluated.Clinical Chemistry, 1988
- Near-Patient Testing: Quality of Laboratory Test Results Obtained by Non-technical Personnel in a Decentralized SettingAmerican Journal of Clinical Pathology, 1988
- Bias and precision of cholesterol analysis by physician's office analyzers.Clinical Chemistry, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Combined Shewhart-cusum control chart for improved quality control in clinical chemistry.Clinical Chemistry, 1977