Abstract
We examined the relationship between analytical sensitivity, precision at the lower limit of the reference interval, and diagnostic performance in hyperthyroidism for one radioimmunoassay and five immunometric assay kits for thyrotropin. The analytical sensitivity of these kits extended from 0.05 to 1.56 milli-int. units/L. Diagnostic efficiencies of the immunometric assays, in discriminating between euthyroidism and hyperthyroidism, ranged between 93% and 98%. There was a highly significant correlation (r = 0.99, P less than 0.001) between analytical sensitivity and diagnostic efficiency. The between-assay coefficients of variations, at the lower limit of the reference interval, ranged from 26% to 87%. There was no correlation (r = 0.36) between precision, at this concentration, and diagnostic efficiency. We conclude that analytical sensitivity and not precision is the major determinant in controlling the diagnostic performance of a thyrotropin assay in hyperthyroidism.