A practical measurement system for the accurate determination of linear thermal expansion coefficients
- 1 August 1984
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 17 (8) , 669-673
- https://doi.org/10.1088/0022-3735/17/8/011
Abstract
An accurate and versatile system for the measurement of linear thermal expansion coefficient is described. The system consists of a double-path optical heterodyne laser interferometer and a parallel spring strip supporting system which provides both versatility and accuracy of measurement. The interferometer has a resolution of 1 nm and the parallel strip supporting system is used to accommodate various shapes and dimensions of specimens to be measured. From the measurements of twelve industrial materials including two standard reference materials at room temperatures, the measurement accuracy is estimated to be within 3 to 7*10-8 K-1 on material having corresponding expansion coefficients of between -1 and 23*10-6 K-1.Keywords
This publication has 3 references indexed in Scilit:
- High-resolution multifold path interferometers for dilatometric measurementsJournal of Physics E: Scientific Instruments, 1983
- A New Optical Interferometer for Absolute Measurement of Linear Displacement in the Subnanometer RangeJapanese Journal of Applied Physics, 1983
- An absolute interferometric dilatometerJournal of Physics E: Scientific Instruments, 1977