Sample loss and its control with internal‐loop injection valves in supercritical fluid chromatography
- 1 September 1989
- journal article
- Published by Wiley in Journal of Microcolumn Separations
- Vol. 1 (5) , 230-233
- https://doi.org/10.1002/mcs.1220010505
Abstract
Large losses of solute can occur with internal‐loop injection valves, especially if the assignments of the injection and waste ports are reversed and a highly volatile diluent is used. However, good performance can be obtained even with the port assignments reversed and with a volatile diluent if a flow restrictor is attached to the waste port.Keywords