The LETR‐Principle: A Novel Method to Assess Electrode‐Tissue Contact in Radiofrequency Ablation
- 1 November 1998
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 9 (11) , 1180-1185
- https://doi.org/10.1111/j.1540-8167.1998.tb00090.x
Abstract
Stable electrode-tissue contact is crucial for successful radiofrequency ablation of cardiac tachyarrhythmias. In this in vitro study, a custom-made radiofrequency generator was used to evaluate the correlation between tip temperature response to a minimal radiofrequency power delivery (Low Energy Temperature Response: LETR-Principle) and electrode-tissue contact as well as lesion size. A battery-powered radiofrequency generator (LETR-Box, 500 kHz, 0.1 to 0.3 W) could measure the temperature increase at the tip electrode with 0.01 degrees C accuracy. The device was tested in vitro using isolated porcine ventricular tissue. For various electrode-tissue settings (i.e., 0 to 0.89 N contact force), the temperature increase (deltaT) due to 0.1-W power delivery for 10 seconds was recorded. Subsequently, for the same electrode-tissue contact, a temperature-controlled radiofrequency ablation was performed (70 degrees C target temperature, 50-W maximum output, 30 sec). Thereafter, the lesion size was measured histologically. To prove the safety of the applied LETR-Principle, the tissue was inspected microscopically after continuous radiofrequency power delivery of 0.3 W for 1 hour with high contact pressure (1.33 N). The delivery of 0.1-W radiofrequency power resulted in an average deltaT of 0.18 degrees +/- 0.13 degrees C. During temperature-controlled radiofrequency ablation, the tip temperature was 59 degrees +/- 8.5 degrees C, resulting in a lesion depth of 4.8+/-0.6 mm. The correlation coefficient between deltaT and contact force was 0.97 and 0.81, respectively, for lesion depth. No lesion was microscopically visible after power delivery of 0.3 W for 1 hour with 1.33 N contact pressure. The LETR-Principle safely indicates electrode-tissue contact and lesion depth under in vitro conditions and can be useful for catheter positioning during radiofrequency ablation procedures.Keywords
This publication has 11 references indexed in Scilit:
- Internal Cardioversion of Atrial Fibrillation with a Single Lead Configuration in a Patient with Massive Diaphragmatic HerniationPacing and Clinical Electrophysiology, 1998
- Temperature Guided Radiofrequency Catheter Ablation of Myocardium: Comparison of Catheter Tip and Tissue Temperatures In VitroPacing and Clinical Electrophysiology, 1997
- Biophysical characteristics of radiofrequency lesion formation in vivo: Dynamics of catheter tip–tissue contact evaluated by intracardiac echocardiographyAmerican Heart Journal, 1997
- Radiofrequency Catheter Ablation of Atrial Flutter That Elicits Inappropriate Implantable Cardioverter Defibrillator DischargePacing and Clinical Electrophysiology, 1997
- Reasons for prolonged or failed attempts at radiofrequency catheter ablation of accessory pathwaysJournal of the American College of Cardiology, 1996
- Temperature Measurement as a determinant of Tissue Heating During Radiofrequency Catheter ablation: An Examination of Electrode Thermistor Positioning for Measurement AccuracyJournal of Cardiovascular Electrophysiology, 1995
- Determinants of Lesion Size During Radiofrequency Catheter Ablation: The Role of Electrode‐Tissue Contact Pressure and Duration of Energy DeliveryJournal of Cardiovascular Electrophysiology, 1991
- Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium.Circulation, 1990
- Control of radiofrequency lesion size by power regulation.Circulation, 1989
- Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free WallPacing and Clinical Electrophysiology, 1989