Effect of surface topology on the osseointegration of implant materials in trabecular bone
- 1 December 1995
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 29 (12) , 1567-1575
- https://doi.org/10.1002/jbm.820291213
Abstract
The importance of surface topology and implant material composition on osseointegration in trabecular bone was investigated using three commercially used implant materials and surface‐texturing procedures which included blasting, high temperature acid etching, and hydroxyapatite (HA) coating. Surface roughness and spacing parameters were measured for each implant group with a laser interferometric profilometer. Cylindrical implants were press‐fit into trabecular bone sites in the knee of mature miniature pigs. After 12 weeks in situ, osseointegration was evaluated by (1) mechanical pushout tests to measure bone‐implant interface strength and (2) quantitative morphometric measurements of the percent implant surface covered by bone. We found that HA‐coated implants showed superior osseointegration in terms of both pushout failure load and surface coverage by bone measurements. An excellent correlation (r2 = .90) was found between the average roughness of the implant surface and pushout failure load. New methods for altering the local topologic and/or chemical state of the implant surface (i.e., by acid etching) may provide an important new avenue of research for improving the osseointegrative properties of orthopedic materials. © 1995 John Wiley & Sons, Inc.Keywords
This publication has 9 references indexed in Scilit:
- The influence of a hydroxyapatite and tricalcium-phosphate coating on bone growth into titanium fiber-metal implants.Journal of Bone and Joint Surgery, 1994
- Histomorphometric assessment of the mechanisms for rapid ingrowth of bone to HA/TCP coated implantsJournal of Biomedical Materials Research, 1993
- Hydroxyapatite ceramic coating for bone implant fixationActa Orthopaedica, 1993
- A review of recent advances in stereology for quantifying neural structureJournal of Neurocytology, 1992
- The Mechanism of Loosening of Cemented Acetabular Components in Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1992
- Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigsJournal of Biomedical Materials Research, 1991
- The effect of surface macrotexture and hydroxylapatite coating on the mechanical strengths and histologic profiles of titanium implant materialsJournal of Biomedical Materials Research, 1987
- The efficiency of systematic sampling in stereology and its prediction*Journal of Microscopy, 1987
- An evaluation of variables influencing implant fixation by direct bone appositionJournal of Biomedical Materials Research, 1985