Thickness of Human Cornea Measured by Topographic Pachometry
- 1 February 1978
- journal article
- Published by Wolters Kluwer Health in Optometry and Vision Science
- Vol. 55 (2) , 97-100
- https://doi.org/10.1097/00006324-197802000-00006
Abstract
In the hands of trained user, the topographic pachometer proves to be an instrument of acceptable clinical precision (standard deviation = 0.006mm). An examination of corneal thinning during waking hours revealed a rate of thinning which was greatest upon lid opening and which declined uniformly throughout the day (p = 0.01). An investigation into menstrually related change in corneal thickness revealed that at least 122 subjects would be required to investigate this change on an acceptable statistical basis (p = 0.05).Keywords