Effect of diluent temperature on creatine kinase values found for lyophilized controls and reference sera.
Open Access
- 1 November 1978
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 24 (11) , 2039-2041
- https://doi.org/10.1093/clinchem/24.11.2039
Abstract
We report the effect of temperature of diluent on creatine kinase activity in several lyophilized controls. Creatine kinase activity was significantly greater when the lyophilized control was reconstituted with diluent at 4 degrees C as compared to 25 degrees C. This is an additional source of variation in creatine kinase controls.This publication has 2 references indexed in Scilit:
- Creatine kinase in serum: 1. Determination of optimum reaction conditions.Clinical Chemistry, 1976