Automatic volumetric capillary cytometry for counting white cells in white cell‐reduced plateletpheresis components
- 1 January 1997
- journal article
- research article
- Published by Wiley in Transfusion
- Vol. 37 (1) , 29-37
- https://doi.org/10.1046/j.1537-2995.1997.37197176948.x
Abstract
BACKGROUND: As the benefits of white cell (WBC)‐reduced blood components become increasingly apparent, the need has arisen for a simple, automated WBC‐counting technique that is sensitive to low WBC concentrations. Automated volumetric capillary cytometry was evaluated for its ability to quantify residual WBCs in WBC‐reduced plateletpheresis components.STUDY DESIGN AND METHODS: The volumetric capillary cytometry system evaluated uses a laser to excite fluorescent dye‐labeled nucleated cells. The number of nucleated cells per μL is reported. Four studies were performed: linearity, precision of results near the value of 5 × 106 WBCs per unit, the limit of detection, and correlation to the Nageotte manual counting method.RESULTS: Assay values correlated to expected values (range, 0–125 WBC/μL) with an r2 > 0.99. In the range of 5 × 106 WBCs per unit the CV was 8.5 percent, and concentration differences of 0.15 log10 were detectable. The limit of detection was 1.0 WBCs per microliter (95% upper confidence limit). The assay correlated to the Nageotte method with an r2 of 0.98, slope of 1.0, and y‐intercept of 2.0 WBCs per μL. Assay results were 10 to 15 percent higher than Nageotte results, in samples with values near 5 × 106 WBCs per unit. Technician time per sample was 2 to 3 minutes.CONCLUSION: Volumetric capillary cytometry is precise and sensitive to small differences in WBC concentration in the range of clinical interest. The device provides an efficient new method for quality assurance and control of WBC‐reduced plateletpheresis products.Keywords
This publication has 20 references indexed in Scilit:
- Volumetric capillary cytometry: A new method for absolute cell enumerationCytometry, 1996
- Cost‐effectiveness of blood transfusion and white cell reduction in elective colorectal surgeryTransfusion, 1995
- Leukodepleted‐ABO‐identical blood components in the treatment of hematologic malignancies: A cost analysisAmerican Journal of Hematology, 1995
- Quantitation of residual white cells in filtered blood components by polymerase chain reaction amplification of HLA DQ‐A DNATransfusion, 1994
- Validation of use of the Nageotte hemocytometer to count low levels of white cells in white cell-reduced platelet componentsTransfusion, 1994
- Identification of the blood component responsible for increased susceptibility to gut‐derived infectionTransfusion, 1993
- Large‐volume hemocytometer chamber for accurate counting of white cells (WBCs) in WBC‐reduced platelets: validation and application for quality control of WBC‐reduced platelets prepared by apheresis and filtrationTransfusion, 1993
- A flow cytometric method for counting very low levels of white cells in blood and blood componentsTransfusion, 1993
- Postoperative infection and natural killer cell function following blood transfusion in patients undergoing elective colorectal surgeryBritish Journal of Surgery, 1992
- Infection or suspected infection after hip replacement surgery with autologous or homologous blood transfusionsTransfusion, 1991