Reconstruction of diagnostic x‐ray spectra by numerical analysis of transmission data
- 1 September 1986
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 13 (5) , 707-710
- https://doi.org/10.1118/1.595834
Abstract
Numerical reconstruction of X-ray spectra from narrow-beam transmission data was tested with published spectra in the range from 45 to 100 kVp. Transmission curves were calculated from the specta to simulate measured data. Spectra were reconstructed from these transmission curves with use of an iterative numerical analysis. Comparison of the calculated spectra with the original spectra shows good agreement, including the tungsten characteristic X-rays. This demonstrates the potential usefulness of measured transmission dta for deducing X-ray spectra in the diagnostic energy range.Keywords
This publication has 4 references indexed in Scilit:
- Determination of x-ray spectra including characteristic line intensities from attenuation dataPhysics in Medicine & Biology, 1984
- Reconstruction of 4-MV bremsstrahlung spectra from measured transmission dataMedical Physics, 1983
- Spectral characterization of 4 MV Bremsstrahlung by attenuation analysisMedical Physics, 1981
- X‐ray spectra vs attenuation data: A theoretical analysisMedical Physics, 1981