Estimation of arterial oxygen and carbon dioxide tensions by a single transcutaneous sensor.
Open Access
- 1 April 1985
- journal article
- research article
- Published by BMJ in Archives of Disease in Childhood
- Vol. 60 (4) , 356-359
- https://doi.org/10.1136/adc.60.4.356
Abstract
A transcutaneous electrochemical sensor designed to estimate arterial oxygen (Pao2) and carbon dioxide (Paco2) tensions simultaneously and continuously was evaluated in newborn infants with respiratory illnesses. After calibration with two dry gas mixtures the sensor gave estimates of Pao2 and Paco2 that were comparable with those obtained by two separate electrodes that are already established in clinical use.This publication has 7 references indexed in Scilit:
- Transcutaneous analysis of arterial PCO2.1982
- A Sensor For Non Invasive Monitoring of Carbon DioxideJournal of Clinical Engineering, 1981
- Effect of Electrode Temperature and in Vivo Calibration on Accuracy of Transcutaneous Estimation of Arterial Oxygen Tension in InfantsPediatrics, 1980
- TRANS-CUTANEOUS ESTIMATION OF ARTERIAL PO2 AND PCO2 IN NEWBORN-INFANTS WITH A SINGLE ELECTROCHEMICAL SENSOR1980
- Single electrochemical sensor for transcutaneous measurement of PO2 and PCO2.1979
- tcPCO2 electrode design, calibration and temperature gradient problems.1978