Closed Circle Systems A New Direction in the Practice of Anesthesia
- 1 October 1980
- journal article
- research article
- Published by Wiley in Acta Anaesthesiologica Scandinavica
- Vol. 24 (5) , 354-361
- https://doi.org/10.1111/j.1399-6576.1980.tb01563.x
Abstract
On the basis of the near-constant uptake of inhalation anesthetics [by humans] at the alveolar membrane, the large closed-circuit volume which acts as a buffer, and the effect of concentration on uptake, a simple and practical technique of closed-circuit anesthesia evolved. After initial high flow wash-in of anesthetic gases into the FRC (functional residual capacity), a nearly constant low fresh-gas flow can be maintained.This publication has 10 references indexed in Scilit:
- Assessment of Vaporizer Performance in Low-Flow and Closed-Circuit AnesthesiaAnesthesia & Analgesia, 1980
- New DirectionsAnesthesiology, 1978
- CLOSED CIRCUIT ANAESTHESIA: A new approachBritish Journal of Anaesthesia, 1977
- Effects of Dry Anesthetic Gases on Tracheobronchial Ciliated EpitheliumAnesthesiology, 1972
- Who is Watching the Patient?Anesthesiology, 1972
- Calibration of Gas Flowmeters with the Bubble BuretteAnesthesiology, 1970
- The Rate of Uptake of Nitrous Oxide in Man1Journal of Clinical Investigation, 1954
- EXPERIMENTAL STUDIES ON CARBON DIOXIDE ABSORBERS FOR ANESTHESIAAnesthesiology, 1941
- Advantages and Technique of Carbon Dioxid Filtration with Inhalation Anesthesia*Anesthesia & Analgesia, 1926