High-Resolution Transmission Electron Microscopy Using Negative Spherical Aberration
- 17 March 2004
- journal article
- Published by Oxford University Press (OUP) in Microscopy and Microanalysis
- Vol. 10 (02) , 174-184
- https://doi.org/10.1017/s1431927604040425
Abstract
A novel imaging mode for high-resolution transmission electron microscopy is described. It is based on the adjustment of a negative value of the spherical aberration CS of the objective lens of a transmission electron microscope equipped with a multipole aberration corrector system. Negative spherical aberration applied together with an overfocus yields high-resolution images with bright-atom contrast. Compared to all kinds of images taken in conventional transmission electron microscopes, where the then unavoidable positive spherical aberration is combined with an underfocus, the contrast is dramatically increased. This effect can only be understood on the basis of a full nonlinear imaging theory. Calculations show that the nonlinear contrast contributions diminish the image contrast relative to the linear image for a positive-CS setting whereas they reinforce the image contrast relative to the linear image for a negative-CS setting. The application of the new mode to the imaging of oxygen in SrTiO3 and YBa2Cu3O7 demonstrates the benefit to materials science investigations. It allows us to image directly, without further image processing, strongly scattering heavy-atom columns together with weakly scattering light-atom columns.Keywords
This publication has 24 references indexed in Scilit:
- Atomic-Resolution Imaging of Oxygen in Perovskite CeramicsScience, 2003
- Investigation of Atomic Displacements at a{111} Twin Boundary inby Means of Phase-Retrieval Electron MicroscopyPhysical Review Letters, 1999
- Dielectric analysis of intergrated ceramic thin film capacitorsIntegrated Ferroelectrics, 1997
- Phase retrieval through focus variation for ultra-resolution in field-emission transmission electron microscopyPhysical Review Letters, 1992
- Direct imaging of the oxygen sublattice in superconductors by high-resolution electron microscopyPhysical Review B, 1991
- Synthesis and properties of the 2:4:7 superconductors(R=Y,Eu,Gd,Dy,Ho,Er)Physical Review B, 1989
- Structure of (CuO)2 double layers in superconducting YBa2Cu3O7Nature, 1988
- Limits on quantitative information from high-resolution electron microscopy of YBa2Cu3O7 superconductorsNature, 1987
- Detection of oxygen ordering in superconducting cupratesNature, 1987
- Oxygen ordering and the orthorhombic-to-tetragonal phase transition inPhysical Review B, 1987