The adverse effect of bromelin in anti-D quantitation using the AutoAnalyzer
- 1 March 1972
- journal article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 25 (3) , 206-209
- https://doi.org/10.1136/jcp.25.3.206
Abstract
With eight of 32 sera significantly higher estimations of anti-D concentration were obtained when test cells premodified with bromelin were used instead of bromelin as a reagent added to the reaction manifold. With one serum the difference in estimation, using the two techniques, revealed a ten-fold discrepancy. The consequences of these findings are discussed in relation to the use of the AutoAnalyzer for the quantitation of anti-D.Keywords