Magnetic field changes in the human brain due to swallowing or speaking
- 1 July 1998
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 40 (1) , 55-60
- https://doi.org/10.1002/mrm.1910400108
Abstract
Variations in the magnetic field in the human brain caused by the processes of swallowing or speaking are measured. In both processes, motion of the pharyngeal muscles, especially the tongue and jaw, alter the susceptibility‐induced magnetic field distribution at the brain slice being imaged. This leads to image warping, compromising the analysis of a time series of images, such as in functional magnetic resonance imaging (fMRI). These dynamic changes are assessed by acquiring a time series of images using a gradient‐echo asymmetric‐spin‐echo sequence (GREASE), a technique in which two images are acquired for each excitation–one during the gradient echo, and one during the latter part of the spin echo. The NMR phase difference between the two images is a measure of the magnetic field distribution. A series of brain images, acquired with this sequence while the subject either swallows or speaks, indicated negative magnetic field changes of up to O.087 ppm in the inferior region of the brain for both speaking and swallowing, and in some speech, additional positive changes of up to O.056 ppm in the frontal region of the brain were indicated.Keywords
This publication has 20 references indexed in Scilit:
- Reduction of physiological fluctuations in fMRI using digital filtersMagnetic Resonance in Medicine, 1996
- Functional magnetic resonance imaging of language cortexInternational Journal of Imaging Systems and Technology, 1995
- Deglutitive aspiration in patients with tracheostomy: Effect of tracheostomy on the duration of vocal cord closureGastroenterology, 1995
- Cerebral cortical activity patterns recorded by functional magnetic-resonance imaging technique during a swallow related motor task in humans: , R. Shaker, I. Lang, A. Jesmanowicz, J.S. Hyde, R.C. Arndorder. MCW Dysphagia Institute, Departments of Medicine, Radiology and Biophysics, Medical College of Wisconsin, Milwaukee, WIGastroenterology, 1995
- A Registration and Interpolation Procedure for Subvoxel Matching of Serially Acquired MR ImagesJournal of Computer Assisted Tomography, 1995
- Artifacts due to stimulus correlated motion in functional imaging of the brainMagnetic Resonance in Medicine, 1994
- MRI-PET Registration with Automated AlgorithmJournal of Computer Assisted Tomography, 1993
- A new two‐dimensional phase unwrapping algorithm for MRI imagesMagnetic Resonance in Medicine, 1992
- Analysis of T2 limitations and off‐resonance effects on spatial resolution and artifacts in echo‐planar imagingMagnetic Resonance in Medicine, 1990