Accuracy of ambulatory blood pressure determination: a comparative study
- 1 January 1991
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 51 (5) , 461-466
- https://doi.org/10.3109/00365519109091640
Abstract
This study was designed to discriminate, according to their accuracy, between three ambulatory pressurometers (Diasys 200R, Novacor; P IV, Del Mar Avionics; SpaceLab 90202, SpaceLab). The evaluation was performed against invasive arterial reference measurements. Accuracy was assessed by calculating the error on pressure (EOP) as the difference between invasive and non-invasive measurement of arterial blood pressure. For the systolic values, accuracy (mean of EOP differences) and uncertainty (SD of these differences) were -0.9 +/- 9.7, -4.3 +/- 10.1 and -16.7 +/- 10.1 mmHg for, respectively, Diasys, PIV and SpaceLab. For diastolic values, they were, respectively, 5.9 +/- 6.7, 6.8 +/- 8.5 and 9.1 +/- 6.6 mmHg. EOP was then separated in two different types of errors: (i) the error of dispersion appreciated by the index of homogeneity calculated by a Lehmann analysis and leading to a statistical classification (ii) the error due to the drift of EOP with the reference value, this last error being easier to correct. Two different behaviours were observed for the EOP: (i) the drift of EOP of systolic values was significantly larger for the oscillometric (SpaceLab) than for the auscultatory (Diasys and P IV) method, with no difference between Diasys and P IV (ii) the homogeneity index was not statistically different among these three devices. These data suggest that, in case the correction of the drift of EOP is carried out, there is no statistical significant difference in accuracy between these three pressurometers. However, in our experimental conditions, the two ambulatory pressurometers recording the Korotkoff sounds have a better accuracy than the one using the oscillometric approach.Keywords
This publication has 20 references indexed in Scilit:
- Ambulatory blood pressure recording to identify hypertensive patients who truly need therapyJournal of Chronic Diseases, 1984
- Left ventricular hypertrophy in patients with hypertension: importance of blood pressure response to regularly recurring stress.Circulation, 1983
- Haemodynamic Alterations Triggered by Blood Pressure Assessments by the DoctorClinical Science, 1982
- Blood Pressure During Normal Daily Activities, Sleep, and ExerciseJAMA, 1982
- CUFF AND AMBULATORY BLOOD PRESSURE IN SUBJECTS WITH ESSENTIAL HYPERTENSIONThe Lancet, 1981
- Evaluation of performance of selected devices for measuring blood pressureThe American Journal of Cardiology, 1973
- Blood Pressure and Risk of Coronary Heart Disease: The Framingham StudyDiseases of the Chest, 1969
- Relationship Between Level of Blood Pressure Measured Casually and by Portable Recorders and Severity of Complications in Essential HypertensionCirculation, 1966
- A SPHYGMOMANOMETER FOR EPIDEMIOLOGISTSThe Lancet, 1964
- An Address ON BLOOD PRESSURE IN MAN: ITS ESTIMATION AND INDICATIONS FOR TREATMENT: Delivered before the Westminster Division of the British Medical AssociationBMJ, 1909