Exhalation flow and pressure-controlled reservoir collection of exhaled nitric oxide for remote and delayed analysis

Abstract
BACKGROUND Expiratory flow rate, soft palate closure, and dead space air may influence exhaled levels of nitric oxide (NO). These factors have not been evaluated in the reservoir collection of NO. METHODS Exhaled NO was collected into a reservoir during a single flow and pressure controlled exhalation. RESULTS NO collected in a reservoir containing silica gel was stable for 24 hours. Nasally delivered 4.8% argon measured by mass spectrometry did not contaminate exhaled argon levels (0.1 (0.02)%) in five volunteers during exhalation against a resistance (10 (0.5) cmH2O), hence proving an effective soft palate closure. Exhaled NO in the reservoir was 11 (0.2) ppb, 8.6 (0.1) ppb, 7.1 (0.6) ppb, and 6.6 (0.4) ppb in five normal subjects and 48.3 (18) ppb, 20.3 (12) ppb, 16.9 (0.3) ppb and 10.1 (0.4) ppb in 10 asthmatic subjects at four studied expiratory flows (5–6, 7–8, 10–11, and 12–13 l/min, respectively), with NO levels equal to direct measurement (7.3 (0.5) ppb and 17.4 (0.5) ppb for normal and asthmatic subjects respectively, pCONCLUSIONS Flow and pressure controlled collection of exhaled NO into a reservoir with silica gel provides values identical to the direct measurement and may be used to monitor asthma at home and where analysers are not on site.