Localized arterial wall drug delivery from a polymer-coated removable metallic stent. Kinetics, distribution, and bioactivity of forskolin.
- 1 August 1994
- journal article
- abstracts
- Published by Wolters Kluwer Health in Circulation
- Vol. 90 (2) , 1003-1011
- https://doi.org/10.1161/01.cir.90.2.1003
Abstract
BACKGROUND: Coronary stenting is associated with two major complications: subacute thrombosis and neointimal proliferation resulting in restenosis. Our hypothesis is that the biocompatibility of metallic stents can be improved by coating with a polymer membrane that delivers agents that favorably modify the local arterial microenvironment. This study evaluates the kinetics, distribution, and bioactivity of the model drug forskolin delivered to the local arterial wall by a polyurethane-coated removable metallic stent. METHODS AND RESULTS: Stents were used in rabbit carotid arteries (n = 20) for as long as 24 hours. The quantity of forskolin bound to the stent decreased exponentially with a half-life of 5.8 hours. Blood concentrations peaked at 140 +/- 39 pg/microL at 4 hours. The adjacent arterial media contained 60 +/- 39 ng/mg, which was 380- and 460-fold greater than the contralateral carotid media and the systemic blood, respectively (P < .0001). Media forskolin concentrations declined exponentially over time with a tissue half-life of 5.0 hours. Drug distributed throughout the vessel wall with decreasing gradients in the radial and axial dimensions consistent with a diffusion process. Removal of the stent was associated with a 100-fold decline in media forskolin concentration within 2 hours. Forskolin release was associated with a sustained 92% increase in carotid blood flow and a 60% decrease in local arterial resistance compared with coated control stents (P < .005). In another set of rabbits (n = 14) using a carotid crush injury, flow-reduction model, forskolin prolonged the time to flow variation and occlusion by 12-fold compared with the use of bare metal stents and 5-fold compared with the use of polyurethane-coated stents (P < .0001). CONCLUSIONS: A polymer-coated metallic stent can deliver forskolin to the local arterial wall in high concentrations relative to the blood or other tissues. High local drug concentrations are dependent on maintaining stent-to-tissue gradients. The delivered drug is biologically active, demonstrating vasodilating and antiplatelet properties.Keywords
This publication has 12 references indexed in Scilit:
- Implantation and recovery of temporary metallic stents in canine coronary arteriesJournal of the American College of Cardiology, 1993
- Multicenter investigation of coronary stenting to treat acute or threatened closure after percutaneous transluminal coronary angioplasty: Clinical and angiographic outcomesJournal of the American College of Cardiology, 1993
- Serial angiographic follow-up after palmaz-schatz stent implantation: Comparison with conventional balloon angioplastyJournal of the American College of Cardiology, 1993
- Surface modification of polymeric biomaterials with poly(ethylene oxide), albumin, and heparin for reduced thrombogenicityJournal of Biomaterials Science, Polymer Edition, 1993
- Novel approach to the analysis of restenosis after the use of three new coronary devicesJournal of the American College of Cardiology, 1992
- Forskolin impregnation of small calibre PTFE grafts lowers early platelet graft sequestration and improves patency in a sheep modelEuropean Journal of Vascular Surgery, 1991
- Cardiovascular Effects of Forskolin (HL 362) in Patients with Idiopathic Congestive Cardiomyopathy—A Comparative Study with Dobutamine and Sodium NitroprussideJournal of Cardiovascular Pharmacology, 1990
- A simple equation for description of solute release I. Fickian and non-fickian release from non-swellable devices in the form of slabs, spheres, cylinders or discsJournal of Controlled Release, 1987
- Effects of forskolin on growth and morphology of cultured glial and cerebrovascular endothelial and smooth muscle cellsInternational Journal of Developmental Neuroscience, 1987