Regional improvement of signal-to-noise and contrast-to-noise ratios in dual-screen CR chest imaging-a phantom study
- 12 June 2001
- journal article
- Published by Wiley in Medical Physics
- Vol. 28 (6) , 1080-1092
- https://doi.org/10.1118/1.1376136
Abstract
The improvement of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in dual-screen computed radiography (CR) has been investigated for various regions in images of an anthropomorphic chest phantom. With the dual-screen CR technique, two image plates are placed in a cassette and exposed together during imaging. The exposed plates are separately scanned to form a front image and a back image, which are then registered and superimposed to form a composite image with improved SNRs and CNRs. The improvement can be optimized by applying specifically selected weighting factors during superimposition. In this study, dual-screen CR images of an anthropomorphic chest phantom were acquired and formed with four different combinations of standard resolution (ST) and high-resolution (HR) screens: ST-ST, ST-HR, HR-ST, and HR-HR. SNRs and their improvements were measured and compared over twelve representative regions-of-interest (ROIs) in these images. A 19.1%-45.7% increase of the SNR was observed, depending on the ROI and screen combination used. The optimal weighting factors were found to vary by only 4.5%-12.4%. Largest improvement was found in the lung field for all screen combinations. Improvement of CNRs was investigated over two ROIs in the lung field using the rib bones as the contrast objects and a 29.2%-43.9% improvement of the CNR was observed. Among the four screen combinations, ST-ST resulted in the most SNR and CNR improvement, followed in order by HR-ST, HR-HR, and ST-HR. The HR-ST combination yielded the lowest spatial variation of the optimal weighting factors with improved SNRs and CNRs close to those of the ST-ST combination.Keywords
This publication has 26 references indexed in Scilit:
- Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.Medical Physics, 2000
- Performance of a 41×41‐cm amorphous silicon flat panel x‐ray detector for radiographic imaging applicationsMedical Physics, 2000
- Full breast digital mammography with an amorphous silicon‐based flat panel detector: Physical characteristics of a clinical prototypeMedical Physics, 2000
- Pixelwise fusion for optimizing SNR in multiple-plate computed radiography imagingIEEE Transactions on Medical Imaging, 1999
- Digital radiology using active matrix readout of amorphous selenium: Theoretical analysis of detective quantum efficiencyMedical Physics, 1997
- Improvement of signal-to-noise and contrast-to-noise ratios in dual-screen computed radiographyMedical Physics, 1997
- Empirical investigation of the signal performance of a high‐resolution, indirect detection, active matrix flat‐panel imager (AMFPI) for fluoroscopic and radiographic operationMedical Physics, 1997
- CCD mosaic technique for large-field digital mammographyIEEE Transactions on Medical Imaging, 1996
- Improved image quality utilizing dual plate computed radiographyThe British Journal of Radiology, 1995
- A clinical comparison between conventional and digital mammography utilizing computed radiographyThe British Journal of Radiology, 1994