Evaluation of the Quantum II and Rapid E identification systems

Abstract
A total of 492 clinical Enterobacteriaceae isolates were tested in the API 20E, Rapid E and Quantum II identification systems. Discrepant identifications among these 3 systems were resolved by repeat testing in the identification systems or use of conventional biochemical tests. Of these isolates, 94.1% were correctly identified with the API 20E and Rapid E systems, and 97.0% were correctly identified with the Quantum II system. An additional 48 non-Enterobacteriaceae isolates were tested with the Quantum II system, and 83.3% were correctly identified. The majority of incorrect identifications with the Rapid E and Quantum II systems were caused by a single aberrant biochemical reaction. Reproducibility of the biochemical reactions obtained with these 2 systems was evaluated by testing 40 organisms in triplicate. Identical biocodes for all three tests were obtained for 10 organisms with the Quantum II system and for 19 organisms with the Rapid E system. Reproducibility of the Quantum II test results was improved with a subsequent modification of the photometer of this system. Both the Rapid E and Quantum II system were inexpensive and were technically easy to inoculate and interpret.