Design and characterization of IndyPET-II:A high-resolution, high-sensitivity dedicated research scanner
- 14 October 2003
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 50 (5) , 1491-1497
- https://doi.org/10.1109/tns.2003.817378
Abstract
The IndyPET-II scanner has been developed as a high-resolution, high-sensitivity research PET scanner for use in small animal, intermediate sized animal, and small field of view human (brain, breast) imaging applications. The system consists of four, approximately plane detector banks mounted on a rotatable gantry. Overall these detectors cover a transaxial field of view (FOV) of 23 cm and an axial FOV of 15 cm. This configuration gives reduced parallax and approximately uniform resolution throughout the FOV. The two pairs of opposing detector banks are offset to give increased sampling density and increased resolution. Characterization measurements using the NEMA-2001 and NEMA-1994 standards are reported. In an axial plane, the averaged (radial and transverse) FWHM resolution is 2.5 mm at the center of the FOV and increases to less than 3.5 mm at the edge of the FOV. The NEMA-2001 sensitivity is 9030 cps/MBq at the center of the scanner and 4250 cps/MBq at r=10cm. The NEMA-1994 sensitivity is 23.0 cps/MBq/ml. Using the NEMA-2001 procedures, the peak true count rate is 95 kcps at an activity concentration of 10.4 kBq/ml and the peak NEC count rate is 31 kcps at an activity concentration of 5.4 kBq/ml. For the NEMA-1994 measurements, the peak true count rate is 130 kcps at a concentration of 18.6 kBq/ml. All measurements were performed using a 350-650 keV energy window.Keywords
This publication has 10 references indexed in Scilit:
- IndyPET-a high resolution, high sensitivity dedicated research scannerPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2003
- Performance evaluation of the microPET P4: a PET system dedicated to animal imagingPhysics in Medicine & Biology, 2001
- Basic imaging performance characteristics of a variable field of view PET camera using quadrant sharing detector designIEEE Transactions on Nuclear Science, 1999
- Performance results of a new DOI detector block for a high resolution PET-LSO research tomograph HRRTIEEE Transactions on Nuclear Science, 1998
- Volume Imaging TomographsPublished by Springer Nature ,1998
- Exact and approximate rebinning algorithms for 3-D PET dataIEEE Transactions on Medical Imaging, 1997
- ECAT ART — a continuously rotating PET camera: Performance characteristics, initial clinical studies, and installation considerations in a nuclear medicine departmentEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997
- Investigation of the performance of the General Electric ADVANCE positron emission tomograph in 3D modeIEEE Transactions on Nuclear Science, 1996
- Investigation of the count rate performance of General Electric Advance positron emission tomographIEEE Transactions on Nuclear Science, 1995
- A comparison of PET detector modules employing rectangular and round photomultiplier tubesIEEE Transactions on Nuclear Science, 1995