Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy
- 1 November 2000
- journal article
- clinical trial
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 175 (5) , 1453-1457
- https://doi.org/10.2214/ajr.175.5.1751453
Abstract
OBJECTIVE. We evaluated the diagnostic accuracy of a grid-controlled fluoroscopy unit compared with a conventional continuous fluoroscopy unit for a variety of abdominal and pelvic fluoroscopic examinations. SUBJECTS AND METHODS. Seventy patients (29 men and 41 women; age range, 24-78 years) were enrolled in one of seven abdominal and pelvic fluoroscopic examinations, including upper gastrointestinal series (n = 20), barium enema (n = 10), voiding cystourethrogram (n = 10), percutaneous abdominal catheter tube injection (n = 10), hysterosalpingogram (n = 10), and percutaneous needle insertion and catheter placement (nephrostomy, percutaneous biliary drainage) (n = 10). Each patient underwent at least 10 sec of continuous fluoroscopy that was randomly and blindly compared with 10-sec periods of pulsed fluoroscopy at 15, 7.5, and 3.75 frames per second. A radiologist outside the examination room, unaware of the frame rate per second, evaluated the procedure in real time on a television monitor. The radiologist assessed image quality and diagnostic acceptability using a scoring system. Statistical analysis was performed using the paired Student's t test. RESULTS. For all procedures at all frame rates, we found no statistically significant superiority of one frame rate over another. For most procedures, the slower frame rates were considered equivalent to continuous fluoroscopy when the images were assessed for image quality and diagnostic confidence. CONCLUSION. Our findings suggest that most abdominal and pelvic fluoroscopic procedures can be performed at substantially lower frame rates than those used for continuous fluoroscopy; adopting this procedure may lead to substantial dose savings for the patient and the fluoroscopy operator.Keywords
This publication has 6 references indexed in Scilit:
- Perceptual comparison of pulsed and continuous fluoroscopyMedical Physics, 1994
- Radiation exposure to patients and medical personnel during radiofrequency catheter ablation for supraventricular tachycardiaThe American Journal of Cardiology, 1992
- Radiation Doses to Patients and Personnel Involved in Embolization of Intracerebral Arteriovenous MalformationsActa Radiologica, 1991
- Effect of pulsed progressive fluoroscopy on reduction of radiation dose in the cardiac catheterization laboratoryJournal of the American College of Cardiology, 1990
- RADIATION REDUCTION BY MEANS OF LOW PULSERATE FLUOROSCOPY DURING CARDIAC CATHETERIZATION AND CORONARY ARTERIOGRAPHYAmerican Journal of Roentgenology, 1974
- Dose Reduction Low Pulse-Rate FluoroscopyRadiology, 1972