Acute and long‐term clinical and angiographic outcome after S‐Stent implantation: S‐Stent multicenter safety and efficacy trial
- 16 July 2004
- journal article
- clinical trial
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 62 (4) , 439-444
- https://doi.org/10.1002/ccd.20066
Abstract
The purpose of this study is to demonstrate safety and effectiveness of the S‐Stent in de novo coronary lesions treated with conventional percutaneous coronary balloon angioplasty. Between January 2000 and June 2001, 120 patients were prospectively enrolled at four study centers. Patients were treated with coronary stenting in a total of 137 lesions. Procedural success was achieved in 100% of 137 attempted lesions. Clinical success was 99.8%. In‐hospital mortality was 0.8%; myocardial infarction occurred in 0.8% and stent thrombosis in 0.8%. After stent implantation, the minimal lumen diameter increased from 0.92 ± 0.43 to 2.74 ± 0.36 mm (P < 0.0001) and the percent diameter stenosis decreased from 68.0 ± 16.2 to 4.5 ± 12.0 (P < 0.0001). At 6‐month follow‐up, the percent diameter stenosis was 33.5 ± 21.3 and the angiographic restenosis rate was 16.5%. Target lesion revascularization was required in 12 patients (10.1%). We conclude that the use of S‐Stent for coronary intervention resulted in a high procedural success rate and low angiographic restenosis at 6 months after implantation. Catheter Cardiovasc Interv 2004;62:439–444.Keywords
This publication has 17 references indexed in Scilit:
- A Clinical Trial Comparing Three Antithrombotic-Drug Regimens after Coronary-Artery StentingNew England Journal of Medicine, 1998
- Diabetes mellitus and the clinical and angiographic outcome after coronary stent placementJournal of the American College of Cardiology, 1998
- Introduction of the multilink stent into routine angioplasty practice: Early angiographic and clinical outcomeCatheterization and Cardiovascular Diagnosis, 1998
- Clinical and Angiographic Outcome of Micro Stent II Implantation in Native Coronary ArteriesThe American Journal of Cardiology, 1998
- First International New Intravascular Rigid-Flex Endovascular Stent Study (FINESS): Clinical and Angiographic Results After Elective and Urgent Stent ImplantationJournal of the American College of Cardiology, 1997
- Coronary stenting with a novel stainless steel balloon-expandable stent: Determinants of neointimal formation and changes in arterial geometry after placement in an atherosclerotic modelJournal of the American College of Cardiology, 1996
- Long-term angiographic and clinical outcome after implantation of a ballon-expandable stent in the native coronary circulationJournal of the American College of Cardiology, 1994
- Emergent use of balloon-expandable coronary artery stenting for failed percutaneous transluminal coronary angioplasty.Circulation, 1992
- Guidelines for percutaneous transluminal coronary angioplastyJournal of the American College of Cardiology, 1988
- Intravascular Stents to Prevent Occlusion and Re-Stenosis after Transluminal AngioplastyNew England Journal of Medicine, 1987