Failure of a Second and Third Generation Implantable Cardioverter Defibrillator to Sense Ventricular Tachycardia: Implications for Fixed-Gain Sensing Devices
- 1 May 1992
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 15 (5) , 749-755
- https://doi.org/10.1111/j.1540-8159.1992.tb06841.x
Abstract
Failure to sense ventricular tachycardia and/or ventricular fibrillation by implantable cardioverter defibrillators (ICDs) is rare. We report a case in which persistent undersensing of monomorphic and polymorphic ventricular tachycardia occurred with a second and third generation ICD using fixed-gain sensing. This occurred despite adequate R wave sensing during sinus rhythm. The use of an endocardial sensing lead did not correct the problem. Failure to sense ventricular tachycardia in the third generation device with fixed-gain sensing occurred late after implantation and was discovered only at follow-up electrophysiology testing of the ICD. This problem could not be corrected by reprogramming of the device, and was not related to lead dislodgement. Placement of a new device with an automatic-gain sensing algorithm and use of previously implanted epicardial leads with better sensing characteristics provided appropriate sensing of ventricular tachyarrhythmias. The case illustrates the importance of testing the sensing of all ventricular arrhythmias in patients with fixed-gain ICD's. Follow-up electrophysiology testing and evaluation of epicardial and endocardial leads may be necessary in certain cases to ensure adequate sensing of ventricular tachyarrhythmias late after implantation.Keywords
This publication has 12 references indexed in Scilit:
- Correlation Between the Ventricular Electrogram Amplitude in Sinus Rhythm and in Ventricular FibrillationPacing and Clinical Electrophysiology, 1990
- Propafenone Induced Acute Variation of Chronic Atrial Pacing Threshold: A Case ReportPacing and Clinical Electrophysiology, 1990
- Implantable cardioverters and defibrillatorsCurrent Problems in Cardiology, 1989
- Effects of Encainide on the Function of Implanted PacemakersPacing and Clinical Electrophysiology, 1989
- Evaluation of antiarrhythmic drugs on defibrillation energy requirements in dogs. Sodium channel block and action potential prolongation.Circulation, 1989
- Implanted Automatic Defibrillators: Effects of Drugs and PacemakersPacing and Clinical Electrophysiology, 1988
- The Automatic Implantable Cardioverter Defibrillator: T Wave Sensing in the Newest GenerationPacing and Clinical Electrophysiology, 1988
- Effect of Antiarrhythmic Drugs on Defibrillation Threshold: Case Report of an Adverse Effect of Mexiletine and Review of the LiteraturePacing and Clinical Electrophysiology, 1988
- Failure of the automatic implantable defibrillator to detect ventricular fibrillationThe American Journal of Cardiology, 1986
- Cardiac pacing and pacemakers III. Sensing the cardiac electrogramAmerican Heart Journal, 1977