The accuracy of esophageal pressure measurements in convalescent and sick intubated infants
- 1 January 1986
- journal article
- research article
- Published by Wiley in Pediatric Pulmonology
- Vol. 2 (1) , 5-8
- https://doi.org/10.1002/ppul.1950020105
Abstract
Assessment of the esophageal pressure technique for measuring dynamic lung compliance (Cdyn) showed that accurate results were obtained in only 5 of 15 studies of sick intubated infants, whereas they were obtained in 6 of 6 studies of convalescent nonintubated infants. In sick infants, inaccuracy was due to large variations in esophageal pressure change (APes) at different esophageal depths and difficulties validating Pes using the occlusion test. Methods of assessing lung stiffness that depend on the measurement of esophageal pressure are unreliable and should not be used in sick intubated infants.Keywords
This publication has 15 references indexed in Scilit:
- Partitioning of respiratory mechanics in halothane-anesthetized humansJournal of Applied Physiology, 1985
- Pulmonary compliance in sick low birthweight infants. How reliable is the measurement of oesophageal pressure?Archives of Disease in Childhood, 1983
- Influence of chest wall distortion on esophageal pressureJournal of Applied Physiology, 1983
- Problems with plethysmographic estimation of lung volume in infants and young childrenJournal of Applied Physiology, 1982
- Failure of body plethysmography in bronchial asthmaJournal of Applied Physiology, 1982
- Measurement of pleural pressure in neonatesJournal of Applied Physiology, 1982
- Critical assessment of jacket plethysmographs for use in young childrenJournal of Applied Physiology, 1982
- Improved esophageal balloon technique for use in infantsJournal of Applied Physiology, 1980
- Nonhomogeneous alveolar pressure swings in the presence of airway closureJournal of Applied Physiology, 1980
- Relationship of Intra-Oesophageal Pressure to Mouth Pressure during the Measurement of Thoracic Gas Volume in the NewbornNeonatology, 1978