Gas Chromatographic Determination of Fumigant Residues in Stored Grains, Using Isooctane Partitioning and Dual Column Packings
- 1 March 1983
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of AOAC INTERNATIONAL
- Vol. 66 (2) , 228-233
- https://doi.org/10.1093/jaoac/66.2.228
Abstract
A gas chromatographic (GO procedure for determining fumigants in grains was developed. Fumigants were leached from grain samples with the official AOAC method using acetone-water (5 + 1). They were then partitioned from the leachate with isooctane, yielding a dry, stable extract that was analyzed by GC. Fortified sample recoveries ranged from 90 to 100%. Two GC columns were used, 20% OV-101 and 20% OV-225/20% OV-17 (2 + 1). These columns gave dissimilar retention profiles and baseline resolution for the 7 fumigants investigated: chloroform, ethylene dichloride, carbon tetrachloride, trichloroethylene, chloropicrin, ethylene dibromide, and tetrachloroethylene. Further tests showed that grain samples could be screened for fumigant residues by direct injection of the acetone-water leachates obtained using the AOAC method.This publication has 1 reference indexed in Scilit: