19 mm Sized Bileaflet Valve Prostheses’ flow Field Investigated by Bidimensional Laser Doppler Anemometry (part II: Maximum Turbulent Shear Stresses)
- 1 November 1997
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 20 (11) , 629-636
- https://doi.org/10.1177/039139889702001106
Abstract
The investigation of the flow field generated by cardiac valve prostheses is a necessary task to gain knowledge on the possible relationship between turbulence-derived stresses and the hemolytic and thrombogenic complications in patients after valve replacement. The study of turbulence flows downstream of cardiac prostheses, in literature, especially concerns large-sized prostheses with a variable flow regime from very low up to 6 L/min. The Food and Drug Administration draft guidance requires the study of the minimum prosthetic size at a high cardiac output to reach the maximum Reynolds number conditions. Within the framework of a national research project regarding the characterization of cardiovascular endoprostheses, an in-depth study of turbulence generated downstream of bileaflet cardiac valves is currently under way at the Laboratory of Biomedical Engineering of the Istituto Superiore di Sanità. Four models of 19 mm bileaflet valve prostheses were used: St Jude Medical HP, Edwards Tekna, Sorin Bicarbon, and CarboMedics. The prostheses were selected for the nominal Tissue Annulus Diameter as reported by manufacturers without any assessment of valve sizing method, and were mounted in aortic position. The aortic geometry was scaled for 19 mm prostheses using angiographic data. The turbulence-derived shear stresses were investigated very close to the valve (0.35 D0), using a bidimensional Laser Doppler anemometry system and applying the Principal Stress Analysis. Results concern typical turbulence quantities during a 50 ms window at peak flow in the systolic phase. Conclusions are drawn regarding the turbulence associated to valve design features, as well as the possible damage to blood constituents.Keywords
This publication has 13 references indexed in Scilit:
- 19 mm Sized Bileaflet Valve Prostheses’ flow Field Investigated by Bidimensional Laser Doppler Anemometry (part I: Velocity Profiles)The International Journal of Artificial Organs, 1997
- Approaching comparability and results of pulsatile flowin vitrotesting of prosthetic heart valvesJournal of Medical Engineering & Technology, 1995
- Determination of Principal Reynolds Stresses in Pulsatile Flows After Elliptical Filtering of Discrete Velocity MeasurementsJournal of Biomechanical Engineering, 1993
- The geometry of the aortic root in health, at valve disease and after valve replacementJournal of Biomechanics, 1990
- Velocity and Shear Stress Distribution Downstream of Mechanical Heart Valves in Pulsatile FlowThe International Journal of Artificial Organs, 1989
- In vitro comparison of aortic heart valve prosthesesThe Journal of Thoracic and Cardiovascular Surgery, 1988
- Turbulent stress measurements downstream of six mechanical aortic valves in a pulsatile flow modelJournal of Biomechanics, 1988
- Two-Component Laser Velocimeter Measurements Downstream of Heart Valve Prostheses in Pulsatile FlowJournal of Biomechanical Engineering, 1986
- In-vitro wall shear measurements at aortic valve prosthesesJournal of Biomechanics, 1984
- Physical effects in red blood cell traumaAIChE Journal, 1969