Reliability and exactness of MRI‐based volumetry: A phantom study
- 1 July 1996
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 6 (4) , 700-704
- https://doi.org/10.1002/jmri.1880060421
Abstract
This study investigated the influence of slice thickness. section orientation, contrast, shape, and sequence type on the exactness of MRI-based volumetry. Ni-doped agarose gel phantoms (4 to 46 ml) were scanned with a T1-weighted three-dimensional Fourier transform (FT) fast low-angle shot (FLASH) and a multiecho two-dimensional FT-Turbo spin-echo (SE) sequence. After segmentation with a three-dimensional region-growing algorithm, the geometric volume was measured considering the partial volume effect. The variability coefficient (Pearson) was .7%. The volumetric error increased with slice thickness, depending on the size and form of the object. Cross sections resulted in smaller error than longitudinal sections (finger-shaped phantoms, nonisotropic image data). Three-dimensional FT was superior to two-dimensional FT imaging. Results of slice thickness and section orientation experiments can be explained by the partial volume effect. Higher errors in two-dimensional FT imaging were caused by object movements between two interleaved acquisitions. The study shows a considerable influence of the imaging parameters on the exactness, which depends on size and form of the structure of interest.Keywords
This publication has 15 references indexed in Scilit:
- Medical workstations for applied imaging and graphics researchComputerized Medical Imaging and Graphics, 1994
- Computer‐assisted identification and quantification of multiple sclerosis lesions in MR imaging volumes in the brainJournal of Magnetic Resonance Imaging, 1994
- Reliability of in vivo volume measures of hippocampus and other brain structures using MRIMagnetic Resonance Imaging, 1993
- Measurement of whole temporal lobe and hippocampus for MR volumetryNeurology, 1993
- Brain Tissue Volume Measurement from Magnetic Resonance Imaging A Phantom StudyInvestigative Radiology, 1993
- Graphical Analysis of MR Feature Space for Measurement of CSF, Gray-Matter, and White-Matter VolumesJournal of Computer Assisted Tomography, 1993
- Magnetic resonance imaging in hereditary and idiopathic ataxiaNeurology, 1993
- Age-Related Differences in Volumes of Subcortical Nuclei, Brain Matter, and Cerebrospinal Fluid in Healthy Men as Measured With Magnetic Resonance ImagingArchives of Neurology, 1992
- MR‐based hippocampal volumetry in the diagnosis of Alzheimer's diseaseNeurology, 1992
- Sources of error in the quantitative analysis of MRI scansMagnetic Resonance Imaging, 1991