Abstract
A special type of phoswich detector system has been evaluated for measurement of radionuclides which decay with emission of time coincident beta and gamma radiation. Background reductions of more than two orders of magnitude have been obtained for the energy region from 500-950 keV. Both NE 102 plastic scintillators and anthracene were evaluated. Advantages and disadvantages of the method are discussed.Keywords
This publication has 2 references indexed in Scilit:
- Evaluation of a Phoswich Detector for the in Situ Analysis of 90SrIEEE Transactions on Nuclear Science, 1980
- Low Background Beta Counting with CaF2(Eu) in a Phoswich ConfigurationIEEE Transactions on Nuclear Science, 1978