Rate Responsive Cardiac Pacing Using a Minute Ventilation Sensor
- 1 August 1993
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (8) , 1650-1655
- https://doi.org/10.1111/j.1540-8159.1993.tb01035.x
Abstract
A minute ventilation sensing rate responsive pacemaker was implanted in 15 patients (8 males and 7 females) with bradycardia. The mean age was 72.8 +/- 8.7 years. The single chamber system measures transthoracic impedance between the tip electrode of a standard bipolar lead and the pulse generator case. In the adaptive mode the pulse generator calculates a rate responsive factor or slope during maximal exercise but functions as in the VVI mode. The patients exercised maximally on an upright cycle ergometer with the pacemaker programmed to VVI mode, adaptive mode, and rate responsive mode. Exercise and gas exchange data were collected continuously and analyzed using an automated breath-by-breath system. The slope, heart rate, and ventilation were measured every 20 seconds. Heart rate in pacemaker dependent patients correlated well to minute ventilation (correlation coefficient ranging from 0.72-0.95, P < 0.0001). This study demonstrates that minute ventilation is a good metabolic sensor in rate responsive pacing.Keywords
This publication has 6 references indexed in Scilit:
- Effects of long-term enalapril therapy on cardiopulmonary exercise performance after myocardial infarction.Circulation, 1991
- Reliability of Minute Ventilation as a Parameter for Rate Responsive PacingPacing and Clinical Electrophysiology, 1989
- Rate Responsive Pacing Using a Minute Ventilation SensorPacing and Clinical Electrophysiology, 1988
- Relationship Between Heart Rate and Minute Ventilation, Tidal Volume and Respiratory Rate During Brief and Low Level ExercisePacing and Clinical Electrophysiology, 1988
- Physiologic correlates of the heart rate response to upright isotonic exercise: Relevance to rate-responsive pacemakersJournal of the American College of Cardiology, 1988